Analog Measurements - Screens and Control Fields
Time
Track/Fixed
Time
This measurement displays the time elapsed from the trigger point to 
the current marker position. 
The 
Ref Set
 function can be used with this measurement to display time 
relative to a specific position. (See 
.) 
Screens on which this field is displayed
SCOPE: all control menus
Time/div 
This field selects the horizontal sweep time per division. 
Operating Considerations
The time-per-division is selected from a list of choices. 
Screens on which this field is displayed
SCOPE: main controls menu
Track/Fixed
When this field is set to 
Track
, the tracking generator is enabled. The 
tracking generator performs a frequency sweep operation. The start 
and stop frequencies are determined by the Span setting (main controls 
menu). The tracking generator allows you to characterize devices (such 
as filter networks) over wide span of frequencies.
An RF offset can be set between the tracking generator and the center 
frequency of the spectrum analyzer. This allows you to look at a signal 
that is related to a source whose frequency is outside the displayed 
span.
Screens on which this field is displayed
SPEC ANL: RF generator controls menu