Chapter 5
157
Analog Measurements - Screens and Control Fields
TX Pwr Zero
TX Pwr Zero
TX Pwr Zero
The transmitter power zero function establishes a 0.0000 W reference 
for measuring RF power at the RF IN/OUT port. 
Operating Considerations
When power is applied to the RF IN/OUT connector, the temperature of 
the internal circuitry increases. This can cause changes in the TX 
power measurement when low power levels are measured immediately 
following high power measurements. 
When alternately making high and low power measurements, always 
zero the power meter immediately before making the low power 
measurements; this provides the best measurement accuracy.
If the 
Auto Zero
 field is set to 
Auto
, the TX power measurement is 
periodically zeroed (automatically). Manual calibration is only needed if 
Auto Zero
 is set to 
Manual
, or if you are measuring highly variable 
signal levels (as mentioned above).
Screens on which this field is displayed
RF ANALYZER
Main Menu