HP Cell Phone Accessories E6380-90019 Manual Do Utilizador

Página de 228
166
Chapter 6 
Configuration - Screens and Control Fields
Duplex Out
Ext Ref In
Duplex Out 
This field is used to indicate losses or gains between the DUPLEX OUT 
port and the device-under-test. 
• Enter a positive value to indicate a gain (such as an amplifier gain). 
The RF generator’s level is automatically set that amount below 
what is indicated in the RF generator’s 
Amplitude
 field. (Example; 
if this value is 10 dB, and the 
Amplitude
 field shows 0 dBm, the 
actual level out this port is 
10 dBm.) The value at the output of the 
external amplifier should then be at the level indicated in the 
Amplitude
 field. 
• Enter a negative value to indicate a loss (such as cable loss). The RF 
generator’s level is automatically set that amount above what is 
indicated in the RF generator’s 
Amplitude
 field to compensate. The 
value at the opposite end of the cable (loss) should then be at the 
level indicated in the 
Amplitude
 field; unless the resulting RF 
generator setting exceeds the maximum output level, then an error 
occurs: 
Input value out of range
. In that case, reduce the 
Amplitude
 setting, or decrease the 
Duplex Out
 value. 
This field is only used when the 
RF Level Offset
 field is set to 
On
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE
Ext Ref In
This field selects the reference frequency of the signal at the EXT REF 
IN connector. You can select 1, 2, 5, or 10 MHz; 1x, 2x, 4x, 8x, or 16x 
chip. 
Operating Considerations
This signal is used as a reference for, and directly affects, these 
functions: 
• RF Generator Frequency (including the Tracking Generator) 
• RF Frequency Counter 
• AF Frequency Counter 
• RF Analyzer Tune Frequency 
• Spectrum Analyzer Center Frequency 
AF generators 1 and 2 are not affected by the external reference; they 
use their own reference. 
Se
 for more information.
Main Menu