Connectors
EVEN SECOND SYNC IN
EXT SCOPE TRIG IN
EVEN SECOND SYNC IN 
This port is an input for the even second clock from the base station 
under test. A positive edge on this connector starts two timers. One 
timer, after reaching terminal counts, starts CDMA frame clocks. The 
other timer starts pilot PN sequence generation. The 
Even Sec In
 
field on the CDMA GENERATOR screen controls this port. See the 
.
EXT REF IN 
This connector allows you to input an external reference. The reference 
frequency is selected using the 
Ext Ref In
 file on the INSTRUMENT 
CONFIGURE screen.
Operating Considerations
When a valid signal is applied to the EXT REF IN port, the Test Set 
automatically switches from internal to external reference if the 
Ref 
Select
 field is set to 
Auto
This signal is used as a reference for, and directly affects, these 
functions: 
RF Generator Frequency (including the Tracking Generator) 
RF Frequency Counter 
AF Frequency Counter 
RF Analyzer Tune Frequency 
Spectrum Analyzer Center Frequency 
AF generators 1 and 2 are not affected using an external reference; they 
use their own reference. 
NOTE
The reference output frequency is always 10 MHz, 
independent of the selected input reference frequency. 
Electrostatic discharges to the EXT REF IN port of 0.5 kV 
or above may cause degradation of performance requiring 
operator intervention.
EXT SCOPE TRIG IN 
This connector provides an external oscilloscope’s trigger input. 
Operating Considerations
When measured with no load on the input, a 5 Volt level is present on 
the connector due to the internal pull-up resistor design. 