Chapter 2 
Operating Overview
Saving and Recalling Instrument Setups
BASE Settings 
The BASE register contains any field settings the user has saved that 
are different from the instrument preset state. It establishes a 
reference point for all future saves. If a base is not saved, the preset 
state is used as the reference.
When you save an instrument setup, the new setup is compared to the 
base settings, and any differences are stored under the register name 
you supply. Because only differences are stored, a much larger number 
of instrument setups can be saved than if the contents of every field was 
saved. 
When you recall an instrument setting, every field is reset to the base 
settings. The saved settings are then used to re-establish the desired 
instrument setup. 
You can define your own base setting. If your desired settings are very 
different from the preset values, you may want to change the BASE 
register. This will decrease the amount of memory used to save each 
setup, and allow you to save many more setups. 
CAUTION
Since each save/recall register only contains the 
differences between the setup being saved and the present 
base register settings, changing the base settings causes 
all other saved setups to be erased from memory 
(including the POWERON setting if one has been saved).
Unless you consistently change the same fields to the 
same value each time you use the instrument, you should 
avoid creating your own BASE settings.
Memory Considerations 
When the 
Save/Recall
 field of the I/O CONFIGURE screen is set to 
Internal
, programs are saved to the same non-volatile RAM used to 
create RAM disk(s) and run IBASIC programs. By saving a large 
number of instrument setups, you reduce the amount of RAM available 
to run programs. If you get a “memory overflow” message while trying 
to load a program, you must clear one or more save/recall registers to 
free RAM space. 
Instrument Hardware Changes 
Recalling a saved register that uses a hardware option that has been 
removed results in unspecified operation. Re-install the needed option 
before attempting to recall the associated register(s). 