Overview of the Test Set’s Built-In Tools
Using RF Tools Program
Discrete Freq Insertion Loss
This test measures the loss of a cable or device at a specified frequency. 
This test is the most accurate method for measuring insertion loss. In 
the menu for this test, you can enter the discrete frequency and signal 
level at which the measurement will be made. 
This test requires two 6-dB pads. The pads are put on the ANT IN and 
DUPLEX OUT ports on the Test Set to reduce impedance mismatch 
error at these ports. A reference level is obtained by connecting a short 
calibration cable between the pads on the DUPLEX OUT and ANT IN 
ports. Next, the test cable or device is added between the calibration 
cable and one of the pads, and a measurement is made of the addition 
loss from the reference level. The insertion loss at the specified 
frequency is displayed on the screen. 
NOTE
The insertion loss at the specified frequency is displayed twice. The 
second display is enlarged for easier reading.
Swept Insertion Loss
This test measures the loss of a cable or device by sweeping it over the 
specified frequency range. In the menu for this test, you can enter the 
frequency range over which the cable is to be swept. 
This test requires two 6-dB pads. The pads are put on the ANT IN and 
DUPLEX OUT ports on the Test Set to reduce impedance mismatch 
error at these ports. A reference level is obtained by connecting a short 
calibration cable between the pads on the DUPLEX OUT an ANT IN 
ports. Next, the test cable or device is added between the calibration 
cable and one of the pads, and a measurement is made of the addition 
loss from the reference level. The trace of the cable loss over the 
frequency range swept is displayed. The average loss is also displayed 
on the screen. 