“Open Air” Calibration Procedure  
for the MW600 Dissolved Oxygen Meter 
 
1.  Take the membrane off the end of the probe and dry the diode off. 
 
2.  Connect the probe to the base unit and turn the unit on.  Your display 
should read 0.0 if it does not -- go to the position “0” screw on the 
front of the unit with the screw driver provided and through the access 
hole make the unit display read 0.0 
 
3.  Go to the membrane and put the “O” ring inside the membrane on the 
shoulder at the base of the threads. Fill the membrane to just short of 
the “O” ring with MA9071 electrolyte solution.  Be sure that there are 
no bubbles inside the membrane. (
 Air bubbles are the main cause of 
calibration failure
 ) 
 
4.  Screw the membrane on to the probe. 
 
5.  Turn the meter on and hang the probe off the end of a table pointing 
down and let the unit display stabilize (this may take 5 to 15 min. ) 
 
6.  Push down and hold the “Cal” button on the front face of the meter 
and with the screwdriver go to the front Cal position “100%” and 
make the display read 100. 
 
You are now calibrated -- Watch for the membrane having wrinkles or 
stretch marks -- if you see these – replace with a new membrane.  Always 
use fresh electrolyte solution model # MA9071 
 
Hope this information is helpful 
 
Milwaukee Instruments Technical Department