Graham Field GF-TX5EMS Manual Do Utilizador

Página de 32
13
TECHNICAL SPECIFICATIONS
Channel
Dual, isolated between channels
Pulse Intensity
Adjustable 0 mA – 100 mA peak into 500 ohm load each channel, 
constant current
Pulse Width
30μs~300μs, adjustable by 10μs/step
Pulse Rate
1Hz-150Hz, 
adjustable
1Hz~20Hz adjustable by 1Hz/step
20Hz~150Hz adjustable by 5Hz/step
Timer
10, 20, 30, 40, 50, 60 minutes and continuous
Patient Compliance 
Counter
To show the usage time by user, counted per each 30 minutes treatment 
time
Patient Lock System
To prevent changes in any set parameter by physician / clinician. To 
unlock or lock the unit, press the UP and DOWN keys simultaneously for 
2-3 seconds
Wave Form
Symmetrical bi-phasic square pulse
LCD
Displays Timer, Function Mode, Pulse Width, Pulse Rate, Cycle ON/
Ramp/OFF Time, Low Battery, Patient Compliance Counter, and Patient 
Lock System
Function Mode
C
Constant Mode
S
Synchronous Mode
A
Alternating Mode
Cycle ON Time 
(Contraction)
Adjustable 1-30 second, 1 second/step
ON Ramp
Adjustable 1-10 second, 1 second/step
Cycle OFF Time 
(Relaxation)
Adjustable 1-30 second, 1 second/step
Voltage
50 volt (peak to peak), at 500 ohm load
Power Source
9-volt battery (alkaline or nickel-cadmium rechargeable)
Battery Life
Approximately 70 hours at normal setting
Note: All electrical specifications are ± 10% 500 ohm load.