Página de 289
 111
Remote Commands
OUTPut Commands
calibration factor (see ) to set the attenuation 
factor to the required value for use as the base 
value for the through-power
Cal
New
= (Through - Power
Base
- Att) + Cal
Current
When you switch the absolute power mode 
OFF
, the last set calibration factor becomes 
active, and the attenuation factor is set so that 
the filter attenuation does not change. That is
Att(dB) = Attenuation
filter
(dB) + Cal(dB)
N O T E
Using any of the 
:INPut:ATTenuation
 commands or queries, or 
any of the 
:INPut:OFFSet
 commands or queries, switches the 
absolute power mode off automatically.
See “:OUTPut:POWer” on page 112 for 
information on setting the through-power.
Switch the mode off (using 
OFF
 or 
0
) to set the 
attenuation of the filter by specifying the 
attenuation and calibration factors. Switch the 
mode on (using 
ON
 or 
1
) to set the attenuation 
of the filter by specifying the through-power.
:OUTPut:APMode? 
Syntax    
:OUTPut:APMode?
Description 
The query returns whether the attenuation of 
the filter is set by the attenuation and 
calibration factors, or by the through-power.
0
 indicates the instrument sets the attenuation 
of the filter from the attenuation and calibration 
factors. 
1
 indicates that the instrument sets the 
attenuation of the filter from the through-
power.