Agilent Technologies 8156A Manual Do Utilizador

Página de 289
122 
Remote Commands
SYSTem Commands
OUTPUT 728;":STAT:QUES:PTR?"
 
ENTER 728;A$
:STATus:PRESet 
Syntax    
:STATus:PRESet
Description    
This command presets all the enable registers 
and transition filters for both the OPERation 
and QUEStionable nodes.
All the bits in the ENABle registers are set to 
0
All the bits in the PTRansition registers are 
set to 1
All the bits in the NTRansition registers are 
set to 0
Example    
OUTPUT 728;":STAT:PRES"
8.8 SYSTem Commands 
:SYSTem:ERRor? 
Syntax    
:SYSTem:ERRor?
Description    
This query returns the next error from the error 
queue (see “The Error Queue” on page 84)
Each error consists of the error code and a short 
description of the error, separated by a comma, 
for example 
0,"No error"
. Error codes are 
numbers in the range -32768 and +32767. 
Negative error numbers are defined by the 
SCPI standard. Positive error numbers are 
device dependent. The errors are listed in 
“Display Messages” on page 275