Agilent Technologies 3A Manual Do Utilizador

Página de 60
2-12
Operation
Using Probes with an Agilent 83491A
To compensate for other devices 
The information in this section applies to both optical and electrical measure-
ments. Since the mainframe’s CAL signal is a voltage source, it cannot be used 
to calibrate to the probe tip when the units are set to Ampere, Watt, or 
Unknown. Instead, set the external gain and external offset to compensate for 
the actual characteristics of the device. If you do not know the actual charac-
teristics, you can refer to the typical specifications that came with the device. 
1
Press the reference receiver module’s front-panel channel 
SETUP key. 
2
Press 
External scale
3
Press 
Atten units RatioAttenuation 1:1, and then Units Ampere (Volt, Watt, or 
Unknown). 
4
Press 
Ext gain, and enter the actual gain characteristics of the device. 
5
Press 
Ext offset, and enter the offset introduced by the device.