GETTING STARTED 
3-15
MON2000
JULY 2010
 Keylock Switch Control
3.6
KEYLOCK SWITCH CONTROL
If your GC unit has the keylock control 
enabled, you will not be able to write changes 
or new data to the GC Controller. The following 
error dialog will display.
To see which Modbus registers are affected by 
the keylock control, view the Communications 
section of the PC Config Report via the Report 
 for instructions and an 
example).
When closed ('on'), the GC keylock switch 
protects many of the GC Controller's 
parameters from being changed using 
MON2000.
Most of the dialog boxes in the Application’s 
menu allows you to view data but does not 
allow you to save changes when the keylock 
switch is closed.  Several items in the Reports 
and Control menus are unavailable when the 
keylock switch is closed.  The Users dialog 
cannot be downloaded to the GC Controller 
when the keylock switch is closed.  A GC 
application cannot be downloaded to the GC 
Controller when the keylock switch is closed. 
Since most of the GC Controller's parameters 
are protected by the keylock switch, it is easiest 
to list the operations that are available when 
the keylock switch is closed.  The following 
operations are available when the keylock 
switch is closed:
The keylock switch affects 
only online operations, i.e. 
those that are performed while 
connected to a GC Controller.
The keylock switch feature is 
only available on controllers 
with BOS version 1.6 and 
later. The GC Controller 
parameters protected by the 
keylock switch may vary 
slightly from version to 
version.