CONTROL FUNCTIONS 
4-7
MON2000
JULY 2010
 Auto BTU Start Up and Valve Timing
4.6
AUTO BTU START UP AND VALVE TIMING
The Auto BTU Start Up process, which takes 
about an hour to complete, automatically 
performs the following sequence of tasks: 
1. Sets the timing for each valve.
2. Matches all the component peaks.
3. Adjusts the timed events based on peak 
integration times.
4. Runs a calibration.
5. Checks the range and order of response 
factors.
6. Adjusts the retention time deviations to 
avoid peak overlapping.
You can also initiate a valve timing run, which 
is the first step of the  Auto BTU Start Up 
process, independently of that process.  See 
more information.
4.6.1 Initiating an Auto BTU Start Up Run
To launch the Auto BTU Start Up process, do 
the following:
1. If the GC is currently performing an 
analysis, select Halt from the Control menu 
to put the GC in Idle mode after the current 
analysis run.