CONTROL FUNCTIONS 
4-15
MON2000
JULY 2010
 Stop Now
provided scroll arrows to increase or 
decrease the highlighted value.
5. Click the 
 button or press the 
enter key to write the selected date/time to 
the GC Controller.
Click the 
 button to abort and 
return to the main window.
4.8
STOP NOW
Use this function to immediately stop all 
analysis runs.
1. Use the Control > Stop Now menu to access 
this function.
2. MON2000 prompts you to confirm your 
selection.
Click the 
 button to proceed.
Click the 
 button to abort and return 
to the main window.
This function forces the system to the Idle mode. If 
Stop Now is performed while an analysis is in 
progress, the components will continue to elute from 
the columns per the Idle mode. All resultant data will 
be lost.
Do not perform a Stop Now unless absolutely 
necessary. Whenever possible, use the Halt function.