 APPLICATION FUNCTIONS
MON2000
Sort Retention Time
JULY 2010
5.2.5 Sort Retention Time
To sort the components listed in the displayed 
CDT by their retention times,
1. Click the 
 button or press the 
F6 
key
 
to access this function.
2. MON2000 automatically sorts the 
components.
5.3
TIMED EVENTS
Use this function to view and/or edit the GC 
Timed Events Tables (TEVs) assigned to and 
used by particular gas streams.
Timed Events Description
• Analysis Time - Input total analysis time. 
LIMITS: 0.0 to 6000.0 seconds
• Cycle Time - Input total cycle time. LIMITS: 
0.0 to 6000.0 seconds
• Off During Baseline (per sample valve) - 
toggle check boxes to select which valve(s) 
remain off during a baseline run
MON2000 sorts 
components by the 
corresponding detector 
number first, then by the 
Retention Time.
Sorting the CDT components by retention time 
may affect one or more of the following:
• Analog Outputs
• Averages
• Bargraphs
• Limit Alarms
• User Defined Calculations
 for details about editing 
from the Chromatogram > Chromatogram 
Viewer menu.
The Cycle Time must be at 
least 10 seconds greater 
than the Analysis Time.