5.2.5 Sort Retention Time
To sort the components listed in the displayed
CDT by their retention times,
1. Click the
button or press the
F6
key
to access this function.
2. MON2000 automatically sorts the
components.
5.3
TIMED EVENTS
Use this function to view and/or edit the GC
Timed Events Tables (TEVs) assigned to and
used by particular gas streams.
Timed Events Description
•
• Analysis Time - Input total analysis time.
LIMITS: 0.0 to 6000.0 seconds
• Cycle Time - Input total cycle time. LIMITS:
0.0 to 6000.0 seconds
• Off During Baseline (per sample valve) -
toggle check boxes to select which valve(s)
remain off during a baseline run
MON2000 sorts
components by the
corresponding detector
number first, then by the
Retention Time.
Sorting the CDT components by retention time
may affect one or more of the following:
• Analog Outputs
• Analog Outputs
• Averages
• Bargraphs
• Limit Alarms
• User Defined Calculations
from the Chromatogram > Chromatogram
Viewer menu.
The Cycle Time must be at
least 10 seconds greater
than the Analysis Time.