APPLICATION FUNCTIONS 
5-33
MON2000
JULY 2010
 Delayed Discrete Alarms
the discrete hardware output to be used for 
alarm purposes.
Output values can range from “1” to “N”, per 
the discrete hardware output identification 
number and the GC application. A discrete 
output of “0 - Not used” indicates that no 
output has been assigned. To configure 
discrete outputs, 
.
5.7.1 Delayed Discrete Alarms
delayed discrete alarm can be set to trigger a 
discrete alarm and activate the appropriate 
discrete output if a previously specified discrete 
input has been closed for a user-defined 
amount of time.
A discrete output can be used to monitor one or 
more inputs.
If you are using discrete outputs to reflect the 
status of GC analysis data alarms, ensure that the 
output assignments set here coordinate with those 
set in the Limit Alarms dialog (
).