Emerson Process Management 3-9000-522 Manual Do Utilizador
5-62
APPLICATION FUNCTIONS
MON2000
FID Configuration
JULY 2010
3. Use the pull down menu to select the
appropriate mode setting (e.g.
AUTOMATIC,
MANUAL,
or
OUT OF SERVICE).
Ensure that
the temperature is constant for the Oven
(i.e. Multivalve System block and column
module kit) and the SSS.
4. Click the OK button to apply the changes or
click the Cancel button to discard your
changes and return to the MON2000 Main
window.
5.15
FID CONFIGURATION
Use this function to view the status
information about the Model 700 Gas Chro-
matograph’s Preamplifier Assembly, as well as
to modify certain FID parameters.
To access the FID Configuration window,
MON2000 must be connected to a GC that has
its serial port #4 configured to use ‘ADC700’.
Column
Function
Factory Default Setting
Column 1
Detector/s or Block Temperature
80
o
C (176.0
o
F)
Column 2
Column Temperature
80
o
C (176.0
o
F)
Column 3
Sample Stream Block Temperature
40
o
C (104.0
o
F)
Column 4
(Spare)
Future Use