 APPLICATION FUNCTIONS
MON2000
FID Configuration
JULY 2010
3. Use the pull down menu to select the 
appropriate mode setting (e.g. 
AUTOMATIC, 
MANUAL, 
or 
OUT OF SERVICE). 
Ensure that 
the temperature is constant for the Oven 
(i.e. Multivalve System block and column 
module kit) and the SSS.
4. Click the OK button to apply the changes or 
click the Cancel button to discard your 
changes and return to the MON2000 Main 
window.
5.15
FID CONFIGURATION
Use this function to view the status 
information about the Model 700 Gas Chro-
matograph’s Preamplifier Assembly, as well as 
to modify certain FID parameters.
To access the FID Configuration window, 
MON2000 must be connected to a GC that has 
its serial port #4 configured to use ‘ADC700’.
Column
Function
Factory Default Setting
Column 1
Detector/s or Block Temperature
80
C (176.0
F)
Column 2
Column Temperature
80
C (176.0
F)
Column 3
Sample Stream Block Temperature
40
C (104.0
F)
Column 4
(Spare)
Future Use