10-4
 MODBUS TEST
MON2000
Set GC Com Parameters via MON2000
JULY 2010
10.2.2 Set GC Com Parameters via MON2000
Set the communication parameters for the GC 
Controller serial port to which your PC, or 
other device, is connected.
To determine or reset the communications 
parameters at the GC Controller,
1. Use the Application > Serial Ports menu
 
to 
access the Serial Ports function. The Serial 
Ports dialog appears.
2. Verify the MON2000 serial port Usage 
setting.
(a) Determine the GC Controller serial port 
to which you are directly connected.
(b) In the Serial Ports dialog, ensure that 
the Usage setting for the port is either 
“SIM-2251” or “User_Modbus”. 
If the setting is incorrect, change it by 
using the provided combo box.
3. Verify the Baud Rate setting. 
WinMB accepts the following baud rates: 
300, 1200, 2400, 4800, 9600, 14400, 19200, 
38400, 57600, 115200.
4. Ensure the remaining serial port 
parameters are correct. Revise as required 
and record for future reference. 
Consult accurate 
documentation of the GC 
installation or visually 
inspect the serial line 
connection at the GC 
Controller Terminal Board.