Emerson Process Management 3-9000-522 Manual Do Utilizador

Página de 516
DATA COMPUTATIONS 
C-5
MON2000
JULY 2010
 Analysis Computations
C.3
ANALYSIS COMPUTATIONS
There are two basic analysis algorithms 
included in the controller:
• Area Analysis – calculates area under 
component peak
• Peak Height Analysis – measures height of 
component peak
C.3.1 CONC. ANALYSIS WITH RESPONSE FACTOR
Calibration
The concentration calculations discussed as 
follows require a unique response factor for 
each component in an analysis. These factors 
may be manually entered by an operator or 
automatically calculated by calibrating the 
system. 
where
Equation C-1  Response Factor Calculation
     or     
ARF
n
area response factor for component “n” in 
area per mole percent (%)
HRF
n
height response factor for component “n”
Area
n
area associated with component “n” in cal-
ibration gas
Ht
n
height associated with component “n” in 
mole percent in calibration gas
Cal
n
amount of component “n” in mole percent 
of calibration gas
standard.
ARF
n
Area
n
Cal
n
---------------
=
HRF
n
Ht
n
Cal
n
-----------
=