C-18
 DATA COMPUTATIONS
MON2000
Indirect Calibration
JULY 2010
The mole% value in the sample gas is then 
calculated as
where
C.4.4 Indirect  Calibration
Component gases  which are NOT found in the 
calibration gas, but may be found in the sample 
gas, can be assigned a response factor which is 
a fixed multiple (the Relative Response Factor 
to a Reference Component) that IS found in the 
calibration gas.
The Relative Response Factors and Reference 
Component Values are included in the 
Component Data Table (following the
Multi-Level Calibration coefficients) 
If the Reference Component is None, then 
normal (direct) calibration is used.
If the Reference Component is defined, (e.g. 
Propane) then the mole% value for the indirect 
component (e.g. neoC5), is calculated as: 
Equation C-23  Mole% Value
P
peak size measured in Sample Gas
Mole % =
aP
3
bP
2
cP
d
+
+
+
responsefactor
-----------------------------------------------
Equation C-24  Mole% Value Indirect Component
mole% (neoC5)  =  mole% (Propane)
P neoC5
(
)
P Propane
(
)
-------------------------------
RRFneoC5
(
)