ANALOG OUTPUT CAL. FOR 2350A 
D-1
MON2000
JULY 2010
 Calibrating by Volts
DA ALOG OUTPUT CAL. FOR 2350A
APPENDIX D, ANALOG OUTPUT CAL. FOR 2350A
The initial analog output adjustment will be set 
at the factory before shipment at standard 
values (4-20 mA or 1-5 V). It may be necessary 
to check and/or adjust these values, depending 
on output cabling or impedance. The 
adjustment may require two persons if the 
units are some distance apart. It will require a 
good digital meter to check the zero and full 
scale values at the receiving end. The scale or 
span value can be adjusted by a PC with 
MON2000 software when the values are known 
at the receiving end. 
It is possible to calibrate the analog outputs 
using different engineering units. Two 
examples are described below using both volts 
and percentages. 
D.1
CALIBRATING BY VOLTS
This example assumes that an analog output 
channel needs to calibrated to a span of 1-5 V. 
To calibrate an analog output by adjusting the 
voltage,
1. Use the Application > Analog Outputs menu 
to access this function. 
2. The Analog Outputs dialog appears. 
The 4-20 mA outputs are 
calibrated similarly except 
the current is measured 
instead of the voltage.