Emerson Process Management 3-9000-522 Manual Do Utilizador
ANALOG OUTPUT CAL. FOR 2350A
D-1
MON2000
JULY 2010
Calibrating by Volts
DA ALOG OUTPUT CAL. FOR 2350A
APPENDIX D, ANALOG OUTPUT CAL. FOR 2350A
The initial analog output adjustment will be set
at the factory before shipment at standard
values (4-20 mA or 1-5 V). It may be necessary
to check and/or adjust these values, depending
on output cabling or impedance. The
adjustment may require two persons if the
units are some distance apart. It will require a
good digital meter to check the zero and full
scale values at the receiving end. The scale or
span value can be adjusted by a PC with
MON2000 software when the values are known
at the receiving end.
It is possible to calibrate the analog outputs
using different engineering units. Two
examples are described below using both volts
and percentages.
D.1
CALIBRATING BY VOLTS
This example assumes that an analog output
channel needs to calibrated to a span of 1-5 V.
To calibrate an analog output by adjusting the
voltage,
1. Use the Application > Analog Outputs menu
to access this function.
2. The Analog Outputs dialog appears.
The 4-20 mA outputs are
calibrated similarly except
the current is measured
instead of the voltage.