UPGRADE 2350A GC S/W AND 2350 EPROMS
MON2000
Standard Applications Prior to v1.50
JULY 2010
E.10.3 Standard Applications Prior to v1.50
For versions of GC Applications prior to version
1.50, the CFG Base Name variable did not
exist. Therefore, a combination of other
variables must be inspected to determine the
standard application that was used to
formulate the current user application.
To determine the standard application used to
create a user application prior to version 1.50,
compare the values of three variables from the
System dialog - Unit Type, Max Archive
Averages, and System Description — with the
values found in
compare the maximum number of streams,
which can be determined from the Streams
option under the Application menu.
There is one more possible complication. If the
variable System Description was edited, it may
not be the same value as described earlier in
. In such an instance, it would be
impossible to distinguish between the
2350_00n (i.e., USASTD1, USASTD2, and
USASTD3) and 2350ISOn (i.e., ISOSTD1,
ISOSTD2, and ISOSTD3) applications.
To overcome this complication, you can inspect
one more variable:
1. Use the Application> User Defined>
Selection menu to access the Selection
dialog.
2. In the Selection dialog, CVTABLE_pri will
be the first variable listed if the application
is one of the ISO applications.