INSTALLATION AND SETUP 
2-43
MON2000
JULY 2010
 Conversion Process
2.13
CONVERSION PROCESS
Use the following process to retrofit a 2350 GC 
Controller to a 2350A GC Controller.
1. At the GC Controller site, remove the 
Controller enclosure's front panel. 
(a) For the explosion-proof Controller, the 
front panel is secured by 16 screws. 
Remove those screws first. 
(b) Then carefully lower the front panel on 
its bottom hinges. The front panel is 
heavy, so make sure it does not drop and 
cause damage. 
(c) For the rack mount Controller, the rear 
of the enclosure is open; it allows access 
for most field wiring procedures without 
removing the enclosure. 
2. Locate the GC Controller's Terminal Board 
for Field Wiring (TB). The TB is attached to 
the GC Controller's card cage assembly, 
facing the enclosure's front panel. (In the 
rack mount Controller, the TB faces 
outward toward the rear of the enclosure.) 
3. Loosen the six screws that secure the TB. 
Then unplug the TB from its connections at 
the back, top of the board. 
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE 
Before removing the unit cover from the GC Controller, make certain the power supply 
switch is OFF and the AC power cord is disconnected. Observe all safety precautions 
when you are working in a hazardous environment. 
Failure to observe all safety precautions could result in serious injury or death.