 INSTALLATION AND SETUP
MON2000
CONNECT VIA GC EXTERNAL MODEM
JULY 2010
2.16
CONNECT VIA GC EXTERNAL MODEM
2.16.1 HARDWARE SETUP
1. Halt any ongoing analysis runs.
Use the Control > Halt menu to access this 
function.
2. Disconnect AC power from the GC 
Controller.
3. For the explosion-proof Controller, remove 
the cover from the GC Controller housing to 
expose the GC Controller Terminal Board 
for field wiring.
For the rack mount Controller, the rear of 
the enclosure is open; it allows access for 
most field wiring procedures without 
removing the enclosure.
Ensure you have a pre-existing PC connection to the GC 
unit before installing the external modem. If the GC unit is 
already on location, do this via another remote operator 
connection, or, onsite, via use of a portable PC and serial 
cable. Ideally, configure the GC unit and install the 
external modem prior to placement at the remote location.
Note: You cannot set the Serial Ports parameters from the 
GC unit front panel.) 
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
Before removing the unit cover from the GC Controller, make certain the 
power supply switch is OFF and the AC power cord is disconnected. Observe 
all safety precautions when you are working in a hazardous environment.
Failure to observe all safety precautions could result in serious injury or death.