Pressure Systems 9816 Manual Do Utilizador
Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©
Page 105
www.PressureSystems.com
Note
To minimize the possibility of overpressure damage, multi-range
9816 modules are manufactured to make use of the CAL (x) port of
the LOWEST installed pressure range.
9816 modules are manufactured to make use of the CAL (x) port of
the LOWEST installed pressure range.
4.3.2. Span
Calibration
Summary
Following is a simple, step-by-step procedure for executing a “full scale” span calibration of a
9816 Intelligent Pressure Scanner. It is assumed that all channels in the unit have the same
full-scale pressure range. Optional commands are shown within brackets [ ].
Description
TCP/IP Data
Insure that valves in RUN/CAL mode (default).
[w1200]
… normal data acquisition
Perform Re-zero calibration
see Sec. 4.2.2
Place the module calibration manifolds into the CAL
position if this is the desired pressure application input.
The RUN position may be a better choice for modules with
transducers having different ranges.
position if this is the desired pressure application input.
The RUN position may be a better choice for modules with
transducers having different ranges.
[w0C01] for CAL pos.
[w0C00] for RUN pos.
[w0C00] for RUN pos.
Apply exact full scale pressure to the appropriate module
CAL and CAL REF inputs [or optionally to RUN inputs]
CAL and CAL REF inputs [or optionally to RUN inputs]
Delay for settling of pneumatic inputs
Verify that measured data reads near expected full scale
[rFFFF0]
Instruct module to calculate new gain coefficients for all 16
channels
channels
ZFFFF
Place calibration manifold back into the RUN position
w0C00
Store new gain coefficients into transducer nonvolatile
memory
memory
w09
… continue normal data acquisition