Pressure Systems 9816 Manual Do Utilizador

Página de 159
Pressure Systems, Inc.                                                                              98RK-1 & 9816 User’s Manual© 
 
 
 
Page 105 
 
www.PressureSystems.com 
 
Note
 
To minimize the possibility of overpressure damage, multi-range 
9816 modules are manufactured to make use of the CAL (x) port of 
the LOWEST installed pressure range.
 
 
4.3.2. Span 
Calibration 
Summary 
 
Following is a simple, step-by-step procedure for executing a “full scale” span calibration of a 
9816 Intelligent Pressure Scanner.  It is assumed that all channels in the unit have the same 
full-scale pressure range.  Optional commands are shown within brackets [ ]. 
 
Description  
TCP/IP Data 
Insure that valves in RUN/CAL mode (default).  
[w1200] 
…  normal data acquisition 
 
Perform Re-zero calibration 
see Sec.  4.2.2 
Place the module calibration manifolds into the CAL 
position if this is the desired pressure application input.  
The RUN position may be a better choice for modules with 
transducers having different ranges. 
[w0C01] for CAL pos. 
 
[w0C00] for RUN pos. 
Apply exact full scale pressure to the appropriate module 
CAL and CAL REF inputs [or optionally to RUN inputs] 
 
Delay for settling of pneumatic inputs 
 
Verify that measured data reads near expected full scale 
[rFFFF0] 
Instruct module to calculate new gain coefficients for all 16 
channels 
ZFFFF 
Place calibration manifold back into the RUN position 
w0C00 
Store new gain coefficients into transducer nonvolatile 
memory 
w09 
… continue normal data acquisition