4.4.2  
Multi-Point Calibration Summary 
 
Following is a simple step-by-step procedure for executing a “multi-point” calibration of a Model 
9816 Intelligent Pressure Scanner
.  It is assumed that all channels in the unit have the same 
full-scale pressure range.  Optional commands are shown within brackets [ ].  If it should 
become necessary to abandon this calibration procedure once it is started, you may execute the 
Abort sub-command [C 03] of ‘C’ at any time after the first ‘C’ sub-command. 
 
Description  
TCP/IP Data 
Insure that valves in RUN/CAL mode (default).  
[w1200] 
…  normal data acquisition assumed to be running 
 
Place the module calibration manifolds into the CAL 
position if this is the desired pressure application input.  
The RUN position may be  a better choice for modules 
with transducers having different ranges. 
[w0C01] for CAL pos. 
 
[w0C00] for RUN pos. 
Ready the module for multi-point calibration by executing 
the Configure & Start (‘00’) sub-command of ‘C’.  This 
establishes all the channels to be affected, and 
determines the total number of calibration points that will 
be supplied (3 in this example) in later steps.  It also 
starts module averaging for calibration (64 samples in this 
example).  The linear fit (1) is required. 
C 00 FFFF 3 1 64 
Apply 1st calibration pressure to the module’s CAL or 
RUN inputs.  The zero (0.0) point is assumed in this case.  
After applying zero pressure verify that this pressure is 
measured correctly by the module. 
[rFFFF0] 
When the data are stable, enter the Collect Data (‘01’) 
sub-command of ‘C’ specifying this first calibration point 
(1) with zero pressure (0.0). 
C 01  1  0.0 
Apply 2nd calibration pressure to the module’s CAL or 
RUN inputs.  A full-scale (+5 psi) point is assumed in this 
case.  Verify that pressure reads correctly. 
[rFFFF0] 
When the data are stable, enter another Collect Data 
(‘01’) sub-command of ‘C’ specifying this second 
calibration point (2) with 5.0 psi pressure. 
 C 01  2  5.0 
Apply 3rd calibration pressure to the module’s CAL or 
RUN inputs.  A mid-scale negative  (-2.5 psi)  point is 
assumed in this case.  A vacuum pump is normally 
required to achieve such a pressure with 903x calibrators. 
Verify that measured pressure reads correctly. 
[rFFFF0] 
When the data are stable, enter last Collect Data (‘01’) 
sub-command of ‘C’ for this point (3) with a negative     
(-2.5 psi) pressure. 
 C 01 3  -2.5