Gas Chromatograph Setup page
Section 5: Startup and meter configuration
 Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
5.6.13
Gas Chromatograph Setup page
Figure 5-16  Field Setup Wizard - Gas Chromatograph Setup page
 The 
Live gas composition from GC
 check box is only available if the optional GC interface feature 
is enabled on the 
Field Setup Wizard - Startup Page
, a Series 100 Option Board (or Series 100 Plus 
Option Board) is installed and a you have a valid 
GCKey (
1.
From the 
Field Setup Wizard - Gas Chromatograph Page
, choose the desired settings:
a.
Serial Port: Select which serial port will be connected to the GC. While the port is 
configured for communications to a GC, it will not act as a Modbus slave device 
for communications from Daniel MeterLink or a SCADA system.
b.
GC protocol: Select the protocol for which the GC is configured. The Mark III uses 
7 data bits, Even parity, and 1 stop bit for ASCII Modbus and 8 data bits, No 
parity, and 1 stop bit for RTU Modbus.
c.
GC baud rate: Select the baud rate for which the GC is configured.
d.
GC comms address: Enter the Modbus ID of the GC.
e.
GC stream number: Enter the stream number for the gas composition the Daniel 
Mark III Gas Ultrasonic meter will read.
f.
GC heating value units: Select the units for which the heating value is configured 
in the GC.
g.
GC heating value type: Select the type of heating value the GC will return.