Philips 1045219 Manual Do Utilizador

Página de 26
BiliTx Parents’ Manual
20
Guidance and Manufacturer’s Declaration - Electromagnetic 
Immunity  
This device is intended for use in the electromagnetic environment specified below. The user of this 
device should make sure it is used in such an environment.
Immunity Test
IEC 60601 Test 
Level
Compliance Level 
Electromagnetic Environment -
Guidance
Electrostatic Discharge 
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete 
or ceramic tile. If floors are covered 
with synthetic material, the relative 
humidity should be at least 30%.
Electrical Fast Transient/
Burst
IEC 61000-4-4
±2 kV for power supply 
lines
±1 kV for input-output 
lines
±2 kV for supply mains
±1 kV for input/output 
lines
Mains power quality should be 
that of a typical home or hospital 
environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV for common mode
Mains power quality should be 
that of a typical home or hospital 
environment.
Voltage dips, short 
interruptions and voltage 
variations on power 
supply input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for 0.5 
cycle 
40% U
T
(60% dip in U
T
) for 5 
cycles
70% U
T
 
(30% dip in U
T
) for 25 
cycles 
<5% U
T
 
(>95% dip in U
T
) for 5 sec
<5% U
T
(>95% dip in U
T
) for 0.5 
cycle 
40% U
T
(50% dip in U
T
) for 5 
cycles
70% U
T
 
(30% dip in U
T
) for 25 
cycles
<5% U
T
 
(>95% dip in U
T
) for 5 sec
Mains power quality should be 
that of a typical home or hospital 
environment. 
Power frequency (50/60 
Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields 
should be at levels characteristic of a 
typical location in a typical hospital 
or home environment.
NOTE: U
T
 is the a.c. mains voltage prior to application of the test level.