6
If therapy is interrupted for any reason, resume therapy as 
 
soon as possible and contact your home care provider.
Carefully place the panel cable to avoid entanglement.
 
Position the Illuminator device on a stable surface, 
 
preferably lower than the infant. When the BiliTx system is 
used with the fiber optic panel configuration and carrying 
case, the device can hang on the outside of a crib or 
treatment area.
Do not place the Illuminator device, power supply, or 
 
carrying case in an incubator or infant warmer. 
Do not place a temperature sensor for the infant warmer or 
 
incubator under the fiber optic panel.
Do not place the Illuminator device, power supply, or 
 
carrying case in a crib or other treatment area next to the 
baby.
Never block the air vents of the unit or place it on a soft 
 
surface such as a bed, crib, carpeted floor, or couch where 
the air vents may be blocked.
During phototherapy, the the baby’s water balance may 
 
become disturbed. Before and during phototherapy, make 
sure the baby is properly hydrated and that his or her body 
temperature is maintained.
After treatment has begun, the baby’s bilirubin level should 
 
be measured to make sure therapy is effective.
The fiber optic panel must not be covered by anything 
 
except with the cover provided. Any other type of cover will 
cause a reduction in light intensity.  The setup instructions 
must be followed exactly.
Cautions
Keep the unit away from any heated surface.
 
Do not scratch, damage, or soil the ferrule end of the panel.  
 
Also, do not place sharp or heavy items on the panel, this 
can damage the panel and affect its light output.
Do not dry the fiber optic panel with artificial heat.
 
If the device is dropped, contact your home care provider.
 
 CAUTION
A caution indicates the 
possibility of damage to the 
device.