Kowa AP-7000 Manual Do Utilizador
<
159
13.
Electromagnetic compatibility
(IEC 60601-1-2)
This instrument is a medical electrical instrument. Medical electrical instruments require special care and are subject to
electromagnetic compatibility (EMC) standards. The following section describes the EMC and precautions regarding
this instrument. When installing or using this instrument, read the description carefully and follow the directions
described.
(The EMC of this instrument was tested based on IEC 60601-1-2.)
1. Please note that portable (or mobile) type radio frequency communication devices (RF communications instrument)
electromagnetic compatibility (EMC) standards. The following section describes the EMC and precautions regarding
this instrument. When installing or using this instrument, read the description carefully and follow the directions
described.
(The EMC of this instrument was tested based on IEC 60601-1-2.)
1. Please note that portable (or mobile) type radio frequency communication devices (RF communications instrument)
may adversely affect this instrument resulting in malfunctioning.
2. The electromagnetic compatibility (EMC) of this instrument was tested with the options and accessories shown
below.
Since using an option or accessory other than those specified may cause malfunctioning of this instrument due to
interferences of external device or cause malfunctioning of external device, use only the options or accessories
specified for this instrument.
Since using an option or accessory other than those specified may cause malfunctioning of this instrument due to
interferences of external device or cause malfunctioning of external device, use only the options or accessories
specified for this instrument.
100V power supply cable:KP300VCTF3X1.25SQKS16A:3m
Personal computer: CE Marking and VCCI approved personal computer
USB cable (Type A―B):2m
3. This instrument is not designed to be used adjacent to an external instrument or placing one on top of another.
Therefore, do not apply such use. Nevertheless, if such use is inevitable, it is necessary to monitor constantly to
make sure the instrument is functioning normally after such use has been adopted.
make sure the instrument is functioning normally after such use has been adopted.
4. Accessories and options used with this instrument are tested for EMC when combined with instrument below.
Since usage with an instrument other than that specified may cause malfunctioning of this instrument due to
interferences of other devices or cause malfunctioning of other devices, do not use any instrument other than
specified below.
Option 1:Personal computer: CE Marking and VCCI approved personal computer
interferences of other devices or cause malfunctioning of other devices, do not use any instrument other than
specified below.
Option 1:Personal computer: CE Marking and VCCI approved personal computer
5. We specified the functions listed in the table blow as the basic performance of this instrument to determine EMC of
this instrument.
Function
Essential performance
Instrument operation functions
Stimulus presentation
Camera image display
Touch panel functions
Display without disturbance