Kowa AP-7000 Manual Do Utilizador

Página de 166
 
 
43 
4.7.2.  Examination parameter setting 
[Common parameters]
 
In  the  description  below,  the  values  provided  with  the  asterisk  mark  ()  are  set  by  default.  You  can  change  those 
parameter values using the parameter value change dialog when necessary.
 
①  Presentation time 
 
Standard: 0.2 sec. 
Optional: You can set a value in the range from 0.1 to 1.2 sec. When you press [Optional], the change button in 
the upper right of the dialog becomes effective. Press this button to open the numeric value input dialog, 
and input a new value for the presentation time. 
 
②  Presentation interval 
 
Standard:  Max. 1.5 sec., Min. 0.6 sec. 
 
Long: 
Max. 3.3 sec., Min. 0.6 sec. 
Auto:  The  maximum  interval  is  automatically  set  according  to  the  response  time  of  the  patient  during  an 
examination. When you press [Auto], Standard and Long appear below it. Then, you can select either one 
for the initial value. 
 
③  Stimulus size 
 
III 
Select from I through V, which correspond to the Goldmann’s stimulus sizes I through V. 
 
④  Blind spot size 
 
II 
Select from I through V, which correspond to the Goldmann’s stimulus sizes I through V. 
 
⑤  Fixation loss 
 
ON 
Select ON to stop the current examination when a fixation loss takes place during the examination. 
Select OFF to perform no fixation loss examination with blind spot stimulation. 
You can re–set the blind spot during an examination only. (P. 79) 
 
⑥  False – positive reaction 
 
ON 
To perform the false – positive reaction examination (ON)/Not to perform (OFF) 
 
⑦  False – negative reaction 
 
ON 
To perform the false – positive reaction examination (ON)/Not to perform (OFF) 
 
⑧  Stimulus color 
 
White 
You can select one from four colors of white, blue, green, and red. 
 
⑨  Background color 
 
White 
You can select one from two colors of white and yellow. 
When the background color is set to yellow, the background intensity becomes 100 cd/m
2
 
⑩  Fixation 
 
Center: Only one point at the center of the dome is set as the fixation target. 
Aux: 4 points at a position of 8.75 degrees from the center of the dome are set as the fixation target. When a 
patient has a visual field defect in the macular area and cannot see the target at the center, select Aux and 
tell the patient to see the center of the 4 points. 
You can set the intensity of the fixation target by moving the slider on the right upward or downward.