Kowa AP-7000 Manual Do Utilizador

Página de 166
 
 
53 
 1 
  Examination method   
A  stimulus  with  the  single  intensity  is  moved  along  the  meridian.  The  points  with  responses  from  the  patient  are 
connected to grasp the shape of the patient’s visual field as an isopter. 
 
 2 
  Parameter setting   
You can set the following parameters: 
False–positive reaction, False–negative reaction, Stimulus color, Background color, Fixation Target, AP–340 
 
 3 
  Isopter setting   
When you select Isopter in the Program Change dialog, [Isopter] in the 
lower left becomes selectable. 
When  you  press  [Isopter],  the  Isopter  dialog  appears  to  allow  you  to 
change  the  parameter  values  dedicated  to  the  isopter  examination 
program. 
 
The  values  provided  with  the  asterisk  mark  ()  are  set  by  default  for 
screening. 
 
①  Test 
 
Auto: Examination is automatically performed according to the other 
set parameters. 
Manual: The location and direction of moving the stimulus can be 
set manually. 
Auto  +  Manual:  Examination  is  performed  automatically  first,  and 
then manually. 
When you select “Manual,” the examination can be conducted on up 
to five isopters and up to 36 directions for each isopter. 
*  You  cannot  select  “Manual”  for  Isopter  +  Screening  1,  Isopter  + 
Screening 2, or Isopter + Threshold. 
 
②  Start 1 
When you select “Periphery 80” or “Center 30,” the movement of the 
stimulus is centripetal (moving from the periphery toward the center). 
 
Periphery 80: Examination starts from the outermost angle in each 
direction. 
Center 30: Examination starts from 30°
 
in each direction. 
* You cannot select “Center 30” for Isopter + Screening 1, Isopter + Screening 2, or Isopter + Threshold. 
 
③  Goal 
 
Center: When there is no response from the patient, the examination in the current direction finishes at the center. 
Center/Goal: When you select “Manual” or “Auto + Manual”, [Center] switches to [Center/Goal]. 
When there is no response from the patient, the examination in the current direction finishes at the center or goal 
depending on whether the direction toward the center has been selected with (Start + Direction) or (Start + 
Goal), respectively. 
Maximum: When there is no response from the patient, the stimulus passes through the center to move to the 
maximum range available for examination, and the examination in the current direction finishes. 
 
④  Start 2 
The default is “OFF.” When you select “Blind spot,” the examination is automatically started from the blind spot after 
the examination specified by Start 1 is completed. 
The parameters set for Start 1 are also applied to Start 2 in the same way. However, only “Stimulus (Size/Intensity)” 
can be set individually for Start 1 and Start 2. 
* You cannot select “Blind spot” for Isopter + Screening 1, Isopter + Screening 2, or Isopter + Threshold. 
⑤  Interval 
 
Standard (12 lines): Meridians on both sides at a distance of 5° from horizontal and vertical meridians, 45° – 225° 
meridians, and 135° – 315° meridians 
45°(8 lines) 
20°(18 lines) 
30°(12 lines)