②  Moving method: “Start + Goal” 
Press [Start + Goal] to set the stimulus moving method. 
Touch the examination start position. Then, select one point (goal) in the moving direction.  If you have selected 
[Center/Goal] for the goal position setting in the Isopter dialog, the stimulus stops at the goal. If you have selected 
[Maximum], the stimulus passes through the goal to move to the maximum range available for the examination until 
the examination in the current direction finishes. (P. 52.) 
To cancel the start point, press [×]. 
 
 
Goal selection screen after start selection 
 
③  Deletion of examination results   
You can delete the results of an examination executed manually during the examination. 
Press [Delete] button and select the examination point you wish to delete. A confirmation message appears. Press 
[OK] to delete the examination result. To cancel the deletion, press [Cancel]. 
※ 
You cannot delete any result of the examination that has been executed automatically. If you wish to carry out the 
examination again at particular points after automatic examination, perform the re-test of each point (P. 122) 
※ 
While the [Delete] button is active, you cannot perform any examination. To continue the examination after deleting 
the examination result, press [Delete] again to deactivate the [Delete] button and proceed with the examination. 
 
 
Confirmation message for examination result deletion