N-EVENT-REPORT
"Workstation" (SCU) can send the following DIMSE services: 
·
N-GET
N-EVENT-REPORT is used by the SCP to report execution status changes to "Workstation" (SCU) in an 
asynchronous way.
N-EVENT-REPORT has the following arguments
:
Table 2.2-33: Notification Event Information 
Event Type Name
Event Type ID
Attribute
Tag
Execution Status Info
(2100, 0030)
Print Job ID
(2100, 0010)
Film Session Label
(2000, 0050)
Pending
1
Printer Name
(2110, 0030)
Execution Status Info
(2100, 0030)
Print Job ID
(2100, 0010)
Film Session Label
(2000, 0050)
Printing
2
Printer Name
(2110, 0030)
Execution Status Info
(2100, 0030)
Print Job ID
(2100, 0010)
Film Session Label
(2000, 0050)
Done
3
Printer Name
(2110, 0030)
Execution Status Info
(2100, 0030)
Print Job ID
(2100, 0010)
Film Session Label
(2000, 0050)
Failure
4
Printer Name
(2110, 0030)
Note:
The SCU only releases the Association after the receipt of the event type Done or Failure, if the print 
job sop class is supported.
"Workstation" (SCU) returns the confirmation from the N-EVENT-REPORT operation. 
N-GET is used to retrieve an instance of the Print Job SOP Class. "Workstation" (SCU) uses the N-GET to 
request the SCP to get a Print Job SOP Instance. "Workstation" specifies the UID of the SOP Instance to be 
retrieved.
Table 2.2-34: Supported N-GET Attributes for a Print Job
Attribute Name
Tag
Value
Execution Status
(2100, 0020)
PENDING, PRINTING, DONE, FAILURE