AGFA HE/001132 Manual Do Utilizador

Página de 93
HE/001132
Page 68 of 92
Document No. 001132, Revision 2.0
Agfa HealthCare
NodeID Livelink : 14140190
14 June, 2007
Table 3.1-1: AE Related Application Profiles, Real World Activities and Roles
Implementation Class UID
1.3.51.0.1.3
Implementation Version Name
3.2
AE Specifications
3.2.1
Offline-Media Application Entity Specification
The Offline-Media Application Entity provides standard conformance to the DICOM Interchange Option of
the Media Storage Service Class. The Application Profiles and roles are listed below:
Table 3.2-1: AE Related Application Profiles, Real World Activities and Roles
Application Profile Supported
Real-World Activity
Roles
SC Option
Portable data for imaging
Export to CD-R
FSC
(creation of a File-Set)
Interchange
3.2.1.1
Real World Activities
3.2.1.1.1
Real World Activity - Export to CD-R
The Offline-Media Application Entity acts as an FSC using the interchange option when requested to export 
SOP Instances from the local database to a CD-R medium.
If the current contents selection does not fit on a single media, an error message is displayed and the export 
will fail.
The user will be prompted to insert an empty CD-R for each export job. The contents of the export job will be 
written together with a corresponding DICOMDIR, which is placed in the root directory, to a single-session 
CD-R. Writing in multi-session mode is supported. The user can cancel an export job in the job queue.
The file names and directory names may reflect the patient name or ID if necessary, but in case the operator 
selects “anonymous export” these names and IDs should not refer to the actual patient. 
The aim of anonymous export is that any data, which could be used to derive the identity of the patient, is 
removed from the image. This includes all dates, UID and other ID’s, patient related information and all 
related physician’s names.
All attributes from the following modules are therefore blanked:
-
Patient
-
Patient Study
-
Patient Medical
-
General Study
-
General Series
-
DX Series
-
SOP Common
-
General Image
 
 
9
XX is the build version number.