Справочник Пользователя для Philips MP60/70/80/90

Скачать
Страница из 414
12 Monitoring Invasive Pressure
Zeroing the Pressure Transducer
152
WARNING
If measuring intracranial pressure (ICP) with a sitting patient, level the transducer with the top of the 
patient’s ear. Incorrect leveling may give incorrect values.
Selecting a Pressure for Monitoring
Tell the monitor which pressure you want to monitor by selecting its pressure label. The label is a 
unique identifier for each type of pressure. When you chose a label, the monitor uses that label’s stored 
color, wave scale and alarm settings.
1
In the 
Setup
 <
Press> 
menu, select 
Label
.
2
Select the appropriate label from the list.
Extended Pressure Label Set
The following additional labels are available if 
Label Set
 is set to 
Full
. This setting can only be 
changed in Configuration Mode. 
Note that if your monitor is connected to an Information Center, the additional labels in the extended 
label set may not be correctly displayed. See the Configuration Guide for your monitor for more 
information. 
Zeroing the Pressure Transducer
To avoid inaccurate pressure readings, the monitor requires a valid zero. Zero the transducer in 
accordance with your hospital policy (at least once per day). You must perform a zero: 
• when you use a new transducer or tubing
• every time you reconnect the transducer cable to the monitor
• if you think the monitor’s pressure readings are not correct. 
Label
Description
Label
Description
ABP
Arterial blood pressure
P
Non-specific pressure label
ART 
Arterial blood pressure (alternative)
PAP
Pulmonary artery pressure
Ao
Aortic pressure
RAP
Right atrial pressure
CVP
Central venous pressure
UAP
Umbilical arterial pressure
ICP
Intracranial pressure
UVP
Umbilical venous pressure
LAP
Left atrial pressure
Label
Description
BAP
Brachial arterial pressure
FAP
Femoral arterial pressure
IC1, IC2
Alternative intracranial pressures 
P1, P2, P3, P4
Alternative non-specific pressure labels