25 Printing Patient Reports
Stopping Reports Printouts
268
Stopping Reports Printouts
To stop Reports printing, in the 
Reports
 menu, select 
Stop Report
 to stop the current print job 
Stop All Reports
 to cancel all queued report printouts
Scheduled Rep.
 to toggle to 
Off
 and switch off scheduled reports.
Setting Up Reports
ECG, Vital Signs, Graphic Trends, and Auto Reports settings must be defined in Configuration 
Mode. As the content of context-linked reports, such as Cardiac Output, Calculations, and Wedge, is 
defined by the content of the procedure window, these reports do not need to be set up.
The content you define in the individual 
Setup Reports
 menus will be used in all reports of this 
type: for example, if you set a print speed of 50 mm/sec in the 
ECG Reports
 menu, every ECG report 
will be printed at this speed, irrespective of format.
Setting Up ECG Reports
The settings you choose in the 
ECG Reports
 menu apply for all ECG reports printed. To set up 
ECG Reports, 
1
Select 
Main Setup
 -> 
Reports
 -> 
ECG Reports
. Note the report layout configured for 
your monitor, either Internat. or Cabrera. This setting is inactive (“grayed-out”) in Monitoring Mode 
and can only be changed in Configuration Mode. 
2
Select 
ECG Gain
 and set the required ECG Gain to define how ECG waves will appear on the 
ECG report printouts. 
3
Select 
Speed
 and select the report print speed.
4
Select 
Annotation
 and toggle to 
On
 if the printed ECG wave should be annotated with beat 
labels or not. See the chapter on ECG for a list of beat labels. Pace pulse marks are automatically 
printed beside the wave for paced patients.
Alarm Limits Report
Graphic and numeric report of all 
current alarm limits
In the 
Alarm Limits
 window, select 
Print 
Limits
Realtime Reports, 
including oxyCRG 
Reports (Neonatal 
Event Review)
patient data and numerics, and either: 
all displayed waves OR all measured 
waves OR all measured RT waves Or all 
measured HiRes waves, OR oxyCRG 
waves (RT waves: ECG Primary lead, 
Pleth, Resp; HiRes waves: btbHR, HiRes 
SpO
2
, HiRes Resp)
Select the 
Realtime Reports
 SmartKey, if 
configured
Loops Report
Individual Loops diagram with 
timestamp, currently measured numerics
In the Loops window, select 
Print Loop
Battery Report
Battery serial numbers, information on 
the currently-measured and potential 
charge status of the battery or batteries
In the 
Battery Status 
window, select 
Print 
Status
.
Report types
Report contents
How to start printing reports