4 Patient Alarms and INOPs
Technical Alarm Messages (INOPs)
76
Ttymp
 INOPS
TEMP
See <Temp label> INOPs (under Temp)
Tven
 INOPS
TEMP
See <Temp label> INOPs (under Temp).
Tvesic
 INOPS
TEMP
See <Temp label> INOPs (under Temp)
UAP
 INOPS
PRESS
See <Pressure label> INOPS (under Pressure).
Unsupported LAN
INOP tone
Monitor
There is a problem with the communication to the network 
and central monitoring is currently not possible. Check the 
connection.If the INOP persists, switch off the monitor and 
contact your service personnel.
User I/F Malfunct
.
INOP tone.
Monitor
Perform a visual and functional check of all the monitor input 
devices. Contact your service personnel.
UVP
 INOPS
PRESS
See <Pressure label> INOPS (under Pressure).
V LEAD OFF
Numeric is replaced by -?- for 10 
seconds; INOP tone.
ECG
The V electrode (IEC: C electrode) has become detached from 
the patient or the lead set has been changed. Reattach the 
electrode or select 
New Lead Setup
 in the 
Setup ECG
 
menu to confirm the new lead set. 
<VueLink option> CHK 
CABLE
INOP tone.
VueLink
No cable or the wrong cable connected to the VueLink module, 
or incorrect device selected. Silencing this INOP switches the 
measurement off. 
VueLink INOP abbreviations may differ slightly depending on 
the device category.
<VueLink option> CHK 
CONF.
INOP tone.
VueLink
The wrong external device has been selected on the VueLink 
module, or the external device has not been correctly setup, or 
the wrong cable has been used to connect the device to the 
VueLink module.
VueLink INOP abbreviations may differ slightly depending on 
the device category.
<VueLink option> CHECK 
SETUP
 
INOP tone.
VueLink
No information was received from the external device. The 
device may be switched off or disconnected.
VueLink INOP abbreviations may differ slightly depending on 
the device category.
VueLnk EQUIP MALF 
INOP tone.
VueLink
Malfunction in the VueLink module. If this message appears 
repeatedly, the module must be replaced. Contact your service 
personnel.
VueLink INOP abbreviations may differ slightly depending on 
the device category.
VueLnk NO CONFIG 
INOP tone.
VueLink
The VueLink module has not been configured during 
installation. The installation process should be completed by 
either your biomedical engineering department or the Philips 
service engineer.
VueLink INOP abbreviations may differ slightly depending on 
the device category.
VueLnk UNPLUGGED
INOP tone.
VueLink
The VueLink module has been unplugged from the rack, or the 
whole rack has been disconnected. The measurement 
automatically disappears from the display. Silencing this INOP 
switches off the measurement.
VueLink INOP abbreviations may differ slightly depending on 
the device category.
INOP Message, Indication
Source
What to do