SECTION 6—OPERATING INSTRUCTIONS
Invacare SOLO
2
18
SECTION 6—OPERATING 
INSTRUCTIONS
Location
 WARNING
NEVER block the air opening of the product or place it on a 
soft surface, such as a bed or couch, where the air opening may 
be blocked. Keep the openings free from lint, hair and the like.
Keep unit at least 3 in (7.6 cm) away from walls, draperies, 
furniture, and the like. 
Locate and position the transportable oxygen concentrator in a well 
ventilated space so that the air intake and the air exhausts are not 
obstructed. 
Recommendations for Optimal Performance
Filters 
Air enters the transportable oxygen concentrator through an air intake filter 
located on the left side of the unit. This filter prevents hair and other large 
particles in the air from entering the unit. Before you operate the 
transportable oxygen concentrator, make sure the filter is clean, dry and 
properly positioned. 
To clean/replace the air intake filter, refer to Cleaning, Care, And 
Operating Temperature:
41°F to 95°F (5°C to 35°C)
Relative Humidity:
15% to 60%
Transport/Storage Temperature:
-2°F to 140°F (-20°C to 60°C)
Allow unit to warm, or cool, to operating 
temperature range before using.
Electrical:
No extension cords. 
Altitude:
Up to 10,000 ft (3046 m) above sea level.
Tubing and Cannula:
7 ft (2.1 m) crush resistant cannula 
(DO NOT pinch).
Environment:
Smoke, pollutant and fume free. No confined 
spaces (example: no closets).
Time of Operation:
Up to 24 hours per day when connected to AC or 
DC power.