AD-4405 Weighing Indicator
 Page 
19 
 
 
5.5.
5.5.
5.5.
5.5.     Gravity Compensation Function
Gravity Compensation Function
Gravity Compensation Function
Gravity Compensation Function    
 
  If the scale is used at the calibration location, it is not necessary to perform this function. 
 
  If there is a difference of gravity acceleration between the installed location and 
calibration location it may cause a weighing error. This function specifies the 
gravity accelerations and corrects the span error. 
 
Note 
  The decimal point is not displayed in the function. Example:   9798   = 9.798 m/s
 
  When span calibration is executed, the gravity acceleration correction will 
be cleared and the two gravity acceleration values will return to the factory 
settings. 
 
Step 1  At the Cal 0 display, press the 
MODE
 key twice, g 5et is displayed and press the 
ENTER
 key to enter the gravity compensation function. 
 
  If you want to cancel the current procedure, press and hold the 
ON/OFF
 key and press 
the 
ESC
 key. Then, no parameters are changed and the calibration mode is finished. 
Press the 
ON/OFF
 key to turn the display off after displaying CanCel. 
 
Step 2  The parameter is displayed with triangle 1. Enter the gravity acceleration of the 
calibration location using the 
0
 - 
9
 keys. The parameter xxxx is the gravity 
acceleration. 
ENTER
 key...The key to store the new gravity acceleration and proceed to step 3. 
ESC
 key .......The key to return to g 5et without changing the value. 
 
Step 3  The parameter is displayed with triangle 2. Enter the gravity acceleration of the 
installed location using the 
0
 - 
9
 keys. The parameter xxxx is the gravity acceleration. 
ENTER
 key...The key to store the new gravity acceleration and proceed to step 4. 
ESC
 key .......The key to return to step 2 without changing the value. 
 
Step 4  Now g xxxx is displayed. Press the 
CAL
 key to store the parameters. Caloff is 
displayed. Proceed to step5. 
 
  Caloff could be displayed by pressing the 
+/-
 key while pressing the 
ON/OFF
 key 
instead of the CAL key. 
 
Step 5  Press the 
ON/OFF
 key to turn the display off. 
 
 
 
 
5.5. Gravity Compensation Function