AD-4405 Weighing Indicator
 Page 
55 
 
15.3.1.  Setting the Interval Time   
Step 1  By pressing the 
SETPOINT(3)
 key while pressing the 
PRINT
 key,   intval  will 
be displayed for 2 seconds. 
 
Step  2  Select which interval time to set using the following keys. The item is displayed.   
 
MODE
 key ....... The key to select serial output   int 5i  or printer  int p  to set. 
 
ESC
 key .......... The key to return to the weight display. 
 
ENTER
 key...... The key to proceed to the time setting (Step 3). 
 
Step  3  Set the interval time using the following keys. The key operation is the same as the 
time settings.   
 
0
 - 
9
 key.......... The key to set the value at the blinking digit, then the blinking 
digit shifts to the right. (Illegal numbers are ignored) . 
 
+/-
 key ............. The key to shift the blinking digit to the right. 
 
CLEAR
 key ...... The key to reset the settings and display the interval time stored 
in memory. 
 
ESC
 key .......... The key to return to Step 2 without changing the interval time 
settings. 
 
ENTER
 key...... The key to store the interval time then return to Step 2. 
 
 
15.3.2.  Interval Data Output / Print output 
 
  Set the data output mode or print mode to “Manual mode” respectively to interval 
data output or print out. 
 
  By pressing the 
START(7)
 key while pressing the 
PRINT
 key the interval timer  
starts and the first data is output or printed. 
 
  At the start, the internal printer prints the interval time setting first. 
 
  By pressing the 
STOP (8)
 key while pressing the 
PRINT
 key the interval timer 
stops and the data at that time is output or printed. The internal printer prints “Intv. 
Stop” before printing the data. 
 
  The internal printer prints unstable data in the interval print mode. 
 
  The internal printer replaces over-range data with spaces and decimal point.   
 
 
 
 
15. Common Printer and Data Output