K O D A K   D i s p l a y   M a n a g e r   S y s t e m   V 4 . 0   U s e r ’ s   G u i d e
     
 3 3
C a l i b r a t i o n
Monitor Evaluation
The first time that you initiate an automatic calibration, the system performs an 
evaluation of your monitor to determine if it can be automatically calibrated. If 
you change monitors, the system runs the evaluation again when you initiate 
automatic calibration. 
To evaluate your monitor:
1. With an enable cable and sensor connected to your system and the Auto 
mode enabled, click Calibrate.
The following window opens.
2. Click Yes
IMPORTANT
You must select Yes and perform this evaluation in order to automatically 
calibrate your monitor.
The following window opens. 
3. Attach the sensor and click OK.
The system reads a series of patches and displays the results.
If your monitor can be automatically calibrated, the following message 
appears. Click OK.