Check & Adjust
5
Check & Adjust
5.1
Overview
Description
Leica instruments are manufactured, assembled and adjusted to the best possible 
quality. Quick temperature changes, shock or stress can cause deviations and 
decrease the instrument accuracy.
It is therefore recommended to check and adjust the instrument from time to time. 
This can be done in the field by running through specific measurement procedures. 
The procedures are guided and have to be followed carefully and precisely as 
described in the following chapters. Some other instrument errors and mechanical 
parts can be adjusted mechanically.
Electronic
adjustment
The following instrument errors can be checked and adjusted electronically: 
l, t
Compensator longitudinal and transversal index errors
i
Vertical index error, related to the standing axis
c
Hz collimation error, also called line of sight error
a
Tilting axis error
ATR
ATR zero point error for Hz and V - option