Справочник Пользователя для IDEAL INDUSTRIES Digital Multimeter

Скачать
Страница из 15
 
Form Number TM61310 
Rev 2 Aug 2007 
Page 12 
Table 4: Ω Resistance Test:   
Step 
Range 
Source 
Reading 
*200 OHM 
2.0Ω
1.6 to 2.4 
*200 OHM 
190.0Ω
188.3 to 191.7 
2K OHM 
1.900K
883 to 1
Ω
1.
.917 
20K OHM 
19.00K
18.83 to 19.17 
Ω
200K OHM 
190.0K
188.3 to 191.7 
Ω
20M OHM 
19.00M
18.40 to 19.60 
Ω
*Lead resistance on the 200 ohm range is not included in error. 
 
Checking the Diode Test Function 
To check the diode test function, do the following: 
1
n the meter. 
.  Connect the calibrator to the VΩ and COM inputs o
2.  Turn the rotary switch to
 
3.  Apply 1.500V DC. The meter display should read approx. 1.500V dc. 
e Function 
 to 
To verify the accuracy of the Temperature function, perform the Following: 
teps 1 - 7 to the VΩTemp and COM inputs using a 
2.  Turn the rotary switch to ºC or ºF (Range is –20 to 750 ºC or –4 to 1382 ºF) 
3.  Compare t
reading
eter dis
 reading in
Note : Meter sele
 the pro
tom
h meas
few seconds. 
4.  If the display reading falls o tside of the range 
wn in step
meter 
does not meet specification. 
Table 5: Temperature Test: 
Reading 
4.  Apply a 20-ohm resistor to the meter, the built-in buzzer buzzes. 
Testing the Temperatur
Before beginning this test allow 30 minutes for the meter and calibrator temperature adapter
reach ambient temperature. 
1.  Apply the temperatures in s
temperature adapter to the meter
he 
 on the m
play to the
 Table 4. 
cts
per range au
atically. Eac
sho
urement takes a 
s 1–7 of Table 4, the 
u
Step 
Range 
Source 
ºC 
-20 ºC 
-30 to -10 
ºC 
0 ºC 
-10 to 10 
ºC 
400 ºC 
391 to 409 
ºC 
750 ºC 
721 to 779 
ºF 
-4 ºF 
-23 to 15 
ºF 
32 ºF 
13 to 51 
ºF 
1382 ºF 
1328 to 1436