Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
83 
To define the REPORT PERIOD, follow the instruction shown in Section 6.7.2.2 then press: 
 
Edit Data Channel Menu
 
 
 
SETUP X.X     
  0) CONC:     ATIMER,     1,          900
 
 
 PREV  NEXT            INS    DEL
   EDIT    
 PRNT 
   EXIT 
 
 
 
SETUP X.X      
NAME:CONC
 
 
    <SET  
  SET> 
 EDIT
 
   PRINT 
 
EXIT
 
Press SET>  key until you reach REPORT PERIOD … 
 
 
 
SETUP X.X     
REPORT PERIODD:DAYS:0
 
    0          0           0   
 
    ENTR    EXIT
 
IIf at any time an illegal entry is selected (e.g., days > 366) 
the ENTR button will disappear from the display. 
 
 
 
SETUP X.X     
REPORT PERIODD:TIME:01:01
 
    0          1           0           0 
 
    ENTR    EXIT
 
 
 
 
SETUP X.X      
REPORT PERIOD:000:01:00
 
 
    <SET    SET>
  EDIT 
   PRINT 
 
EXIT
 
Exits to the main 
Data Acquisition
 
menu. 
ENTR
 accepts the new string and 
returns to the previous menu. 
EXIT
 ignores the new string and 
returns to the previous menu. 
Press buttons to set hours 
between reports in the format : 
HH:MM (max: 23:59). This is a 
24 hour clock . PM hours are 13 
thru 23, midnight is 00:00. 
Example 2:15 PM = 14:15
 
Set  the number of days 
between reports (0-366).
 
Use the PREV and NEXT 
buttons to scroll to the 
data channel to be edited. 
From the DATA ACQUISITION menu 
(see Section 6.7.2.2)