Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
86 
6.7.2.11. Disabling/Enabling Data Channels 
Data channels can be temporarily disabled, which can reduce the read/write wear on the disk-on-chip.  
The ALL_01 channel of the GFC 7000TA, for example, is disabled by default.   
To disable a data channel, follow the instruction shown in Section 6.7.2.2 then press: 
 
 
 
 
 
Edit Data Channel Menu
 
 
 
SETUP X.X     
  0) CONC:     ATIMER,     1,          900
 
 
 PREV  NEXT            INS    DEL
   EDIT    
 PRNT 
    EXIT 
Exits to the main 
Data Acquisition 
menu 
 
 
 
SETUP X.X      
NAME:CONC
 
 
    <SET  
  SET> 
 EDIT
 
   PRINT 
 
EXIT
 
 
 
 
SETUP X.X      
CHANNEL ENABLE:ON
 
 
    <SET    SET>
  EDIT 
   PRINT 
 
EXIT
 
ENTR
accepts the new 
setting and returns to the 
previous menu. 
EXIT
 ignores  the new  setting 
and returns to the  previous 
menu. 
SETUP X.X        
CHANNEL ENABLE:ON
 
  
 OFF   
    
 
 
 
    ENTR    EXIT
 
 
Toggle to turn 
channel ON or OFF 
Press SET>  key until… 
From the DATA ACQUISITION menu 
(see Section 6.7.2.2)