Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
106 
6.11.5.  COM Port Communication Modes 
Each of the analyzer’s serial ports can be configured to operate in a number of different modes, listed in 
Table 6-10 which can be combined by adding the mode ID numbers.  For example, quiet mode, computer 
mode and internet-enabled mode would carry a combined mode ID of 11, the standard configuration on 
the GFC 7000TA COM2 port.  Note that each COM port needs to be configured independently. 
Table 6-10:   Com Port Communication Modes 
MODE
1
 ID 
DESCRIPTION 
QUIET 
Quiet mode suppresses any feedback from the analyzer (DAS reports, and warning 
messages) to the remote device and is typically used when the port is communicating 
with a computer program such as APICOM.  Such feedback is still available but a 
command must be issued to receive them. 
COMPUTER  
Computer mode inhibits echoing of typed characters and is used when the port is 
communicating with a computer program, such as APICOM. 
SECURITY  
When enabled, the serial port requires a password before it will respond.  The only 
command that is active is the help screen (? CR).   
HESSEN 
PROTOCOL 
16 
The Hessen communications protocol is used in some European countries.  Teledyne 
Instruments part number 02252 contains more information on this protocol. 
E, 7, 1 
2048 
When turned on this mode switches the com port settings 
 from 
No parity; 8 data bits; 1 stop bit 
to 
Even parity; 7 data bits; 1 stop bit 
RS-485 
1024 
Configures the COM2 Port for RS-485 communication.  RS-485 mode has precedence 
over multidrop mode if both are enabled. 
MULTIDROP 
PROTOCOL 
32 
Multidrop protocol allows a multi-instrument configuration on a single communications 
channel.  Multidrop requires the use of instrument IDs. 
ENABLE 
MODEM 
64 
Enables to send a modem initialization string at power-up.  Asserts certain lines in the 
RS-232 port to enable the modem to communicate. 
ERROR 
CHECKING
128 
Fixes certain types of parity errors at certain Hessen protocol installations. 
XON/XOFF 
HANDSHAKE
2
 
256 
Disables XON/XOFF data flow control also known as software handshaking. 
HARDWARE 
HANDSHAKE 
Enables CTS/RTS style hardwired transmission handshaking.  This style of data 
transmission handshaking is commonly used with modems or terminal emulation 
protocols as well as by Teledyne Instrument’s APICOM software. 
HARDWARE 
FIFO
2
 
512 
Improves data transfer rate when on of the com ports. 
COMMAND 
PROMPT 
4096 
Enables a command prompt when in terminal mode. 
1
 Modes are listed in the order in which they appear in the  
SETUP
  MORE  com  COM[1 OR 2]  MODE menu 
2
 The default sting for this feature is ON.  Do not disable unless instructed to by Teledyne Instruments Customer Service 
personnel.