Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
144 
The pin assignments for the Status Outputs are: 
Table 6-24:   Status Output Pin Assignments 
OUTPUT # 
Status Definition 
Condition 
SYSTEM OK 
On if no faults are present. 
CONC VALID 
On if CO
2
 concentration measurement is valid.   
If the CO
2
 concentration measurement is invalid, this bit is OFF.   
HIGH RANGE 
On if unit is in high range of DUAL or AUTO range modes. 
ZERO CAL 
On whenever the instruments zero point is being calibrated. 
SPAN CAL 
On whenever the instruments span point is being calibrated. 
DIAG MODE 
On whenever the instrument is in diagnostic mode. 
7  
ALARM1 
On whenever the measured CO
2
 concentration is above the set point for 
ALM1
 
ALARM2 
On whenever the measured CO
2
 concentration is above the set point for 
ALM2
 
EMITTER BUS 
The emitters of the transistors on pins 1-8 are bused together. 
DC POWER 
+ 5 VDC 
 
 
Digital Ground 
The ground level from the analyzer’s internal DC power supplies.