Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
158 
6.15.4.7. Setting Hessen Protocol Status Flags 
Teledyne Instruments’ implementation of Hessen protocols includes a set of status bits that the 
instrument includes in responses to inform the host computer of its condition.  Each bit can be assigned 
to one operational and warning message flag.  The default settings for these bit/flags are: 
Table 6-31:   Default Hessen Status Bit Assignments 
STATUS FLAG NAME 
DEFAULT BIT 
ASSIGNMENT 
WARNING FLAGS 
SAMPLE FLOW WARNING 
0001 
BENCH TEMP WARNING 
0002 
SOURCE WARNING 
0004 
BOX TEMP WARNING 
0008 
WHEEL TEMP WARNING 
0010 
SAMPLE TEMP WARNING 
0020 
SAMPLE PRESSURE WARNING 
0040 
INVALID CONC 
(The Instrument’s Front Panel Display Will Show The 
Concentration As “XXXX”) 
0080 
OPERATIONAL FLAGS
 
Instrument Off 
0100 
In Manual Calibration Mode 
0200 
In Zero Calibration Mode 
0400 
In Span Calibration Mode 
0800 
UNITS OF MEASURE FLAGS
 
UGM 0000 
MGM 2000 
PPB 4000 
PPM 6000 
SPARE/UNUSED BITS 
100, 1000, 8000 
UNASSIGNED FLAGS (0000)
 
Sync Warning 
Relay Board Warning 
Conc Alarm 1 
Analog Cal Warning 
Conc Alarm 2 
Cannot Dyn Zero 
Photo Temp Warning 
Cannot Dyn Span 
System Reset 
Invalid Conc 
Rear Board Not Detected 
 
 
NOTES: 
It  is  possible  to  assign  more  than  one  flag  to  the  same  Hessen  status  bit.    This 
allows the grouping of similar flags, such as all temperature warnings, under the 
same status bit.   
Be careful not to assign conflicting flags to the same bit as each status bit will be 
triggered if any of the assigned flags is active.