Calibration Procedures 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
173 
7.5.2. Use of Zero/Span Valves with Remote Contact Closure 
Contact closures for controlling calibration are located on the rear panel CONTROL IN connector.  
Instructions for setup and use of these contacts are found in Section 6.15.1.2.  When the contacts are 
closed for at least 5 seconds, the instrument switches into zero or span mode.  The remote calibration 
contact closures may be activated in any order.  It is recommended that contact closures remain closed 
for at least 10 minutes to establish a reliable reading. 
The instrument will stay in the selected mode for as long as the contacts remain closed.  If calibration is 
enabled, the GFC 7000TA will re-calibrate when the contact is opened, then go into SAMPLE mode.  If 
calibration is disabled, the instrument will return to SAMPLE mode, leaving the calibration unchanged. 
7.6. Automatic Zero/Span Cal/Check (AutoCal) 
The AutoCal system allows unattended periodic operation of the ZERO/SPAN valve options by using the 
GFC 7000TA’s internal time of day clock.  AutoCal operates by executing SEQUENCES programmed by 
the user to initiate the various calibration modes of the analyzer and open and close valves appropriately.  
It is possible to program and run up to 3 separate sequences (SEQ1SEQ2 and SEQ3).  Each sequence 
can operate in one of 3 Modes, or be disabled. 
Table 7-1:     AUTOCAL Modes 
MODE NAME 
ACTION 
DISABLED 
Disables the Sequence 
ZERO 
Causes the Sequence to perform a zero calibration/check 
ZERO-SPAN 
Causes the Sequence to perform a zero and span concentration 
calibration/check 
SPAN 
Causes the Sequence to perform a span concentration 
calibration/check 
 
For each mode there are seven parameters that control operational details of the SEQUENCE.  They are: 
Table 7-2:     AutoCal ATTRIBUTE Setup Parameters 
ATTRIBUTE NAME 
ACTION 
TIMER ENABLED 
Turns on the Sequence timer 
STARTING DATE 
Sequence will operate after Starting Date 
STARTING TIME 
Time of day sequence will run 
DELTA DAYS 
Number of days to skip between each Seq.  execution 
DELTA TIME 
Number of hours later each “Delta Days” Seq is to be run 
DURATION 
Number of minutes the sequence operates 
CALIBRATE 
Enable to do a calibration – Disable to do a cal check only